Front Panel Controls

FRONT PANEL CONTROLS AND COMPONENTS USED TO DISPLAY THE WAVEFORM

The CRT DISPLAY SCREEN is used to display the signal (the figure below). It allows you to make accurate measurements using the vertical and horizontal graticules, as discussed earlier.


CRT display and graticule.


COMPONENTS USED TO ADJUST CRT DISPLAY QUALITY

The front panel controls in the figure below allow you to adjust for a clear signal display. They also allow you to adjust the display position and magnify the horizontal trace by a factor of 10 (X10). Keep in mind that the controls may be labeled differently from one model to another, depending on the manu-facturer. Refer to the figure below as you study the front panel controls descriptions in the next paragraphs.

Quality adjustment for CRT display.


INTEN (Intensity) Control

The INTEN (intensity) control (sometimes called BRIGHTNESS) adjusts the brightness of the beam on the CRT. The control is rotated in a clockwise direction to increase the intensity of the beam and should be adjusted to a minimum brightness level that is comfortable for viewing.

FOCUS and ASTIG (Astigmatism) Controls

The FOCUS control adjusts the beam size. The ASTIG (astigmatism) control adjusts the beam shape. The FOCUS and ASTIG controls are adjusted together to produce a small, clearly defined circular dot. When displaying a line trace, you will use these same controls to produce a well-defined line. In the figure below, view A, shows an out-of-focus beam dot. View B shows the beam in focus. Views C and D show out-of-focus and in-focus traces, respectively.

Effects of FOCUS and ASTIG (astigmatism) control.


TRACE ROTATION

Control The TRACE ROTATION control (figure 2 above) allows for minor adjustments of the horizontal portion of the trace so that you can align it with the horizontal lines on the graticule.

BEAM FINDER Control

Occasionally, the trace will actually be located off the CRT (up or down or to the left or right) because of the orientation of the deflection plates. When pushed, the BEAM FINDER (figure 2) pulls the beam onto the screen so that you can use the horizontal and vertical POSITION controls to center the spot.

Horizontal and Vertical POSITION Controls

The horizontal and vertical POSITION controls (figure 2) are used to position the trace. Because the graticule is often drawn to represent a graph, some oscilloscopes have the positioning controls labeled to correspond to the X and Y axes of the graph. The X axis represents horizontal movement; the Y axis represents the vertical movement. The figure below shows the effects of positioning controls on the trace.

Effects of horizontal and vertical controls.


In view A, the horizontal control has been adjusted to move the trace too far to the right; in view B, the trace has been moved too far to the left. In view C, the vertical POSITION control (discussed later) has been adjusted to move the trace too close to the top; in view D, the trace has been moved too close to the bottom. View E (the figure above) shows the trace properly positioned.

10X MAG (Magnifier) Switch

The 10X MAG (magnifier) switch (figure 2 above) allows you to magnify the displayed signal by a factor of 10 in the horizontal direction. This ability is important when you need to expand the signal to evaluate it carefully.

As you can see, there are many different front panel controls and knowing the function of each will help you better understand how the oscilloscope is used.

(back) (top) (next) (return to oscilloscope page)